User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2527 Verification Procedure (VER1.0) DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 39 NUMBER OF LINES: 121 CONFIGURATION: Fluke 5700A ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM P/N: @ * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Auto Calibration * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ B 1.002 ASK- J F W 1.003 MMFC C2 1.004 HEAD { Safety! } 1.005 DISP Check the safety of the PM 2527 multimeter according 1.005 DISP to the general testprocedure. 1.006 HEAD { Initial Process } 1.007 DISP ! Plug in the model PM 2527 and turn it on. 1.007 DISP ! Before performing the Verification test 1.007 DISP ! allow the unit 30 minutes to warm up. 1.008 HEAD { Zero Test } 1.009 DISP Place a shorting strap between "0" and "V" of the UUT. 1.009 DISP Select the V-DC function and the 2V range (manual). 1.009 DISP Select "AUT"omatic with the "START" switch. 1.010 EVAL Does the {UUT read 0.0000V ñ 1 Digit ?} 2.001 DISP Remove the shorting strap. 2.002 HEAD 2.003 DISP ![27][91]1m Connect as follows 2.003 DISP ! ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 2.003 DISP ![27][91]1m 5700 UUT 2.003 DISP ! Output HI ÄÄÄÄÄÄÄÄÄÄÄ> Volt Input 2.003 DISP ! Output LO ÄÄÄÄÄÄÄÄÄÄÄ> Common (0) 2.004 HEAD { DC Voltage Test } 2.005 DISP Select Autorange and the Volt DC function. 2.006 5700 190mV 0.08U 2W 3.001 5700 0.3V 0.0005U 2W 4.001 5700 1.00V 0.0006U 2W 5.001 5700 1.90V 0.0008U 2W 6.001 5700 -1.90V 0.0008U 2W 7.001 5700 19.0V 0.008U 2W 8.001 5700 190V 0.08U 2W 9.001 5700 900V 0.6U 2W 10.001 HEAD { AC Voltage Test, (60Hz range) } 10.002 DISP Select Autorange and the Volt AC function. 10.003 5700 19mV 0.08U 60H O 2W 11.001 5700 190mV 0.8U 60H 2W 12.001 5700 0.3V 0.005U 60H 2W 13.001 5700 1.00V 0.006U 60H 2W 14.001 5700 1.90V 0.008U 60H 2W 15.001 HEAD { AC Voltage Test, (1kHz range) } 15.002 5700 19.0V 0.08U 1kH 2W 16.001 HEAD { AC Voltage Test, (100kHz range) } 16.002 5700 19.0V 0.12U 100kH 2W 17.001 HEAD { AC Voltage Test, (1kHz range) } 17.002 5700 190V 0.8U 1kH 2W 18.001 5700 600V 5.0U 1kH 2W 19.001 HEAD { Ohms Test } 19.002 DISP ![27][91]1m Connect as follows 19.002 DISP ! ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 19.002 DISP ![27][91]1m 5700 UUT 19.002 DISP ! Sense HI ÄÄÄÄÄÄÄÄÄ¿ 19.002 DISP ! Output HI ÄÄÄÄÄÄÄÄÄÁÄ> Ohm Input 19.002 DISP ! Output LO ÄÄÄÄÄÄÄÄÄÂÄ> Common (0) 19.002 DISP ! Sense LO ÄÄÄÄÄÄÄÄÄÙ 19.003 DISP Select Autorange and the Ohms function. 19.004 5700 0.19kZ 0.00020U CW 20.001 5700 1.9kZ 0.0014U CW 21.001 5700 19kZ 0.014U 2W 22.001 5700 190kZ 0.14U 2W 23.001 5700 1.9MZ 0.0029U 2W 24.001 5700 19MZ 0.029U 2W 25.001 5700 100MZ 0.7U 2W 26.001 HEAD 26.002 DISP ![27][91]1m Connect as follows 26.002 DISP ! ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 26.002 DISP ![27][91]1m 5700 UUT 26.002 DISP ! Output LO ÄÄÄÄÄÄÄÄÄÄÄ> Common (0) 26.002 DISP ! AUX Out ÄÄÄÄÄÄÄÄÄÄÄ> A Input 26.003 HEAD { DC Current Test } 26.004 DISP Select Autorange and the DC Amps function. 26.005 5700 1.9uA 0.0029U 2W 27.001 5700 19uA 0.029U 2W 28.001 5700 190uA 0.29U 2W 29.001 5700 1.9mA 0.0029U 2W 30.001 5700 19mA 0.029U 2W 31.001 5700 190mA 0.29U 2W 32.001 5700 1900mA 2.9U 2W 33.001 HEAD { AC Current Test, (60Hz range) } 33.002 DISP Select Autorange and the AC Amps function. 33.003 5700 19uA 0.09U 60H 2W 34.001 5700 190uA 0.9U 60H 2W 35.001 5700 1.9mA 0.009U 60H 2W 36.001 5700 19mA 0.09U 60H 2W 37.001 HEAD { AC Current Test, (1kHz range) } 37.002 5700 190mA 0.9U 1kH 2W 38.001 5700 1900mA 9U 1kH 2W 39.001 HEAD 39.002 DISP ! P R O C E D U R E C O M P L E T E 39.002 DISP ! Disconnect all cables 39.002 DISP ! [27][91]5m Complete required paperwork